Medical Equipment Surgical Infusion Pump Patient Monitor OSP-500
1. 4.3'' color segment LCD screen, backlight display, can be used in various lighting conditions
2. Simultaneous display: Time, Battery indication, Injection state, Mode, Speed, Injection volume and time, Syringe size, Alarm sound, Block, Accuracy, Body weight, Drug dose and liquid amount
3. Speed, time, volume and drug amount can be adjusted via remote control, easier operation, save time of doctor and nurse
4. Advanced technology, based on Linux system, more safe and steady
5. Multi injection modes: Volume/Time/Body weight mode
6. Visible and audible alarms cover all abnormal conditions
Parameters:
Injection Rate Range
50ml syringe : 0.1~999.9ml/h (step :0.1ml/h)
1000~1800ml/h (step : 1ml/h)
30ml syringe :0.1~900ml/h (step :0.1ml/h)
20ml syringe :0.1~600ml/h (step :0.1ml/h)
10ml syringe :0.1~300ml/h (step :0.1ml/h)
5ml syringe :0.1~150ml/h (step :0.1ml/h)
Purge Rate
50ml syringe: 1800ml/h (Bolus rate: 1200ml/h)
30ml syringe:900ml/h (Bolus rate: 600ml/h)
20ml syringe:600ml/h (Bolus rate: 400ml/h)
10ml syringe:300ml/h (Bolus rate: 200ml/h)
5 ml syringe:150ml/h (Bolus rate: 100ml/h)
Occlusion Alarm Threshold
High: 800mmHg±200mmHg (106.7kPa± 26.7kPa)
Medium: 500mmHg ± 100mmHg (66.7kPa± 13.3kPa)
Low: 300mmHg ± 100mmHg (40.7kPa± 13.3kPa)
Specifications:
Model |
Medical Infusion Pump OSP-500 |
LCD Display |
4.3 inch |
Mechanical Precision |
Within±2% |
Preset Volume Range |
0.1~1999.9ml |
KVO Rate |
0.1~5ml/h (step: 0.1 ml/h) |
Fuse |
F1AL/250V,pcs |
Power Supply |
100~220V, 50Hz |
Size |
282mm(L)×210mm(W)×148mm(H) |
Weight |
2.2Kg |
Environment Requirement:
Operation: a) Ambient temperature: +5ºC~+40ºC
b) Relative humidity: 20%~ 90%
c) Atmosphere Pressure: 86.0~106.0 (kPa)
Transportation & Storage: a) Ambient temperature: -30ºC~+55ºC
b) Relative humidity: ≤95%
c) Atmosphere Pressure: 50.1~106.0 (kPa)
Alarm
Near, Injection finished, Preset dose injection finished, Occlusion, Syringe off, Syringe dislocated, Low battery, AC power off