• Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s
  • Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s
  • Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s
  • Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s
  • Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s
  • Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s

Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s

Customized: Non-Customized
Certification: CE, ISO13485
Colour: Blue
Type of Body Fluid-Processing Device: Human Blood Processor
Dimensions: 1380 X 380 X 400 mm
Gross Weight: 88 Kg
Customization:
Diamond Member Since 2014

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company, Group Corporation

Basic Info.

Model NO.
Hemodialysis ME2800S
Connected Load
<2000va
Pressure Range
1.5-6.0 Bar
Water Inlet Temperature
5ºC~32ºC
Max. Drain Height
1m
Temperature/Flow
30 to 60ºC / 300 to 800ml/Min
Selectable
100-800 Ml/Min
Sensitivity
0.35 Ml Blood/Min (Hct=32)
Transport Package
Carton
Specification
CE ISO
Trademark
Mermaid
Origin
China

Product Description

Dialysis Portable Medical Therapy Dialysis Hemodialysis ME2800S 

Main performance of this device:
1. Self-checking function;
2. Carbonate dialysis;
3. Double needle dialysis;
4. Liquid level detector;
5. Bubble detector;
6. Blood leakage detector;
7. Temperature and electric conductivity monitoring;
8. Arterial pressure, venous pressure and transmembrane pressure monitoring;
9. Rolling blood pump;
10. Heparin pump;
11. Dehydration amount controlled by capacity;
12. Automatic disinfection cleaning program;
13. Stand-by power of blood pump in case of power failure;
14. Information display function of display screen.
Technical Data
General data:
Dimensions                 1380 x 380 x 400 mm (H x W x D);
Weight approx.              88 KG;

Water supply:
Water inlet pressure           1.5-6.0 bar
Water inlet temperature        5ºC~32ºC
Max. drain height             1m   

Concentrate supply
Supply pressure              0 to 2 m suction height

Electrical data
Power supply                AC220V±10%, 47-63HZ
Current consumption max.  max. 10 A

Extracorporeal circuit
Arterial pressure monitoring
Display range               -380 mmHg to +400 mmHg
Accuracy                   ±10 mmHg
Resolution                  1 mmHg

Venous pressure monitoring
Display range               -180 mmHg to +600 mmHg
Accuracy                   ±10 mmHg
Resolution                  1 mmHg

Transmembrane pressure monitoring
Display range               -180 mmHg to +600 mmHg
Accuracy                   ±20 mmHg
Resolution                  1 mmHg   


Arterial blood pump
Blood flow range            30 to 600ml/min in 8 mm bloodline systems
Accuracy                   ±10%          
Air bubble pump
By ultrasound transmission, additional optical monitoring in venous clamp.
Heparin pump:
Delivery range            0ml/h to 10ml/h
Bolus function max.        20 mL per bolus
Syringe size              10, 20, 30, 50ml

Dialysis fluid circuit
Dialysis fluid flow range    
Selectable                100-800 mL/min

Dialysis fluid temperature
Selectable                34.0ºC to 40.0ºC

Dialysis fluid conductivity
Range                   12 to 16 mS/cm (25ºC)     
Accuracy                 ±0.1 mS/cm

Acid concentration dialysis fluid
Default mixing ratio       1+34 (others possible)
Range                  125 to 150 mmol/L

Bicarbonate concentration dialysis fluid
Default mixing ratio       1+27.6 (others possible)
Range                  -8 to +8 mmol/L bicarbonate

Ultrafiltration
UF rate                    0ml/h~4000ml/h
Accuracy                  ±1%
Allowed dialyzer UF factor    unlimited
Parameter displayed          UF goal, UF time, UF rate, UF volume

Blood leak detector
Sensitivity                 ≤0.35 mL blood/min (HCT=32)
KT/V 
Calculation

Disinfection and cleaning programs
Rinse
Temperature/flow       30 to 60ºC / 300 to 800ml/min

Hot Rinse (recirculation)
Temperature/flow   30 to 85ºC / 300 to 800ml/min

Hot disinfection with chemical
Temperature/flow   30 to 85ºC/ 300 to 800ml/min

Various program combination selectable

Technical Data-Options
Blood pressure Monitor (BPM) (optional function)
Display range   Systole:   40-280 mmHg
              Diastole:  40-280 mmHg
Accuracy: 1 mmHg

Central Delivery system for acid conc.
Supply pressure   0 to 100 mbar; 1 m suction height with Central delivery system: 0-500 mbar

Endotoxin filter-- Dialysis fluid filter system  (optional function)
Balancing accuracy   ±0.1% of dialysate flow
Dialysis Portable Medical Therapy Dialysis Hemodialysis Me2800sDialysis Portable Medical Therapy Dialysis Hemodialysis Me2800sDialysis Portable Medical Therapy Dialysis Hemodialysis Me2800sDialysis Portable Medical Therapy Dialysis Hemodialysis Me2800sDialysis Portable Medical Therapy Dialysis Hemodialysis Me2800sDialysis Portable Medical Therapy Dialysis Hemodialysis Me2800s








 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Diamond Member Since 2014

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company, Group Corporation
Management System Certification
ISO 9001, ISO 13485, FSC
Terms of Payment
LC, T/T