How To Use The Infant Incubator BIN3000BS:
First Point: Preparation
1. Do a good job of cleaning the whole machine and disinfection.
2. Lock the foot wheel of the whole machine to prevent the whole machine from moving.
3. The power cord -the end is inserted from the power supply input jack from the side of the chassis, and the other end is correctly inserted into the 220V/50Hz power socket, which is reliable.
4. Add an appropriate amount of distilled water to the water cup.
Second Point: Operation method
1. Turn on the power switch → the temperature controller enters the self -inspection status. Except for the ultra -temperature alarm indicator and the power -off alarm indicator, all indicators and digital tubes are on → All lights are destroyed → Display the temperature of the box and set the temperature value → If you do not press any keys, automatically enter the box temperature control status → The temperature value of the boot setting is to keep the setting value of the last shutdown.
2. Press the settings → Enter the box temperature setting status → set the temperature display window flickering. According to clinical needs, the temperature setting temperature is changed, push the add and subtract key to make the setting temperature display window consistent with the required value → Exit the setting state → enter the box temperature control state (if you do not press the box temperature key, wait for about 15 seconds to automatically enter the box temperature control status.
Third Point: Nursing method
When your baby cares, open the front panel lock, put down the front door panel, you can put in the baby, close the front door panel and lock the lock. When the baby needs to care, you can use the elbow to press the door lock and reach into the elliptical door for operation. In order to avoid the baby's cold, reduce the door opening time as much as possible. Hurry up and close the door after the completion.
Function of Infant Incubator
The main components of the infant incubator are composed of hosting (including baby cabin, chassis, control box, infusion rack and tray), sensor box, skin temperature sensor, seat seat (cabinet or lifting cabinet), radiation light box (jaundice treatment device) (optional accessories), and weighing device (optional accessories).
The infant incubator has a temperature control system and a humidity control system. The temperature control system includes two control working modes: air mode and skin temperature mode. Control of temperature heating output under air mode: The air temperature detected by the temperature sensor in the infant incubator is from the thermal output of the control heater; the control of the temperature heating output under the skin temperature mode: The skin temperature detected according to the temperature sensor comes from the heat output of the control heater. Humidity control system: The humidity detected by the humidity sensor in the infant incubator comes from the humidifier in the water tank.
Infant Care Incubator BIN3000BS parameters | |
Packing size | 108*64*112cm |
N.W. | 92Kg/72Kg |
Mattress temp. uniformity | ≤0.8ºC |
Environment relative humidity | ≤80% |
Atmospheric Pressure | 500~1060hpa |
Skin mode alarm | ±1.0ºC |
Skin sensor accuracy | ≤±0.3ºC |
Air control accuracy | ≤0.5ºC |
Phototherapy unit (option) |
The Precautions Of Medical Infant Incubator:
1. The infant incubator is a high -risk medical device that endangers the safety of human life. It can only be used in similar wards such as neonatal ward, pediatric ward, obstetric ward, and children's intensive care unit. The operator of the infant incubator must be trained and operates under the guidance of a practicing physician.
2. When working in the instrument, the operator must always pay attention to the patient's condition, and regularly monitor and record the temperature of the patient to see if there are abnormal conditions such as overheating or over -cooling. It is recommended to monitor the baby temperature at least once at least once.
3. When any abnormalities and faults occur in the equipment, stop using it immediately, turn off the power supply and transfer patients, and timely notify the company's after -sales service department or the maintenance provider authorized by the company. Do not repair and test by non -authorized personnel.
4. Direct sunlight or other radiant heat sources will increase the temperature of the infant incubator, but it will not start the ultra -temperature alarm function of the infant incubator. Therefore, the infant incubator cannot be placed in a place where the sunlight or other radiant thermal sources exist.
5. Do not place the infant incubator in places where flammable and chemical gases or other flammable substances (such as certain cleaning liquids) exist.
6. Do not place the culture box in places produced by strong electromagnetic fields. Portable and mobile radio frequency communication devices may affect this device.
7. Device that is susceptible to magnetic field interference shall not be placed near the infant incubator, and the infant incubator may interfere with them.
Product feature:
Air temp.and skin temp. servo-controlled by computer>37ºC temperature setting
Various self-check alarms
Free-step mattress tilting adjustment
Removable humidity reservoir , easy to clean
With the transfusion shelf and tray
Added with guardrail
RS-232 connectorPhototherapy(option)